Development, Validation and Green Analytical Assessment of a Stability-Indicating RP-HPLC Method for Simultaneous Estimation of Glycopyrronium, Vilanterol and Fluticasone Furoate in Capsule Dosage Form
K.E. Pravallika*, Kothapalli Narendra, P. Ravi
University College of Pharmaceutical Sciences, Acharya Nagarjuna University, Guntur, Andhra Pradesh, India.
*Corresponding Author E-mail: elvina2108@gmail.com
ABSTRACT:
For simultaneous evaluation of glycopyrronium, vilanterol, and fluticasone furoate in capsules, an Inertsil ODS C18 column was used to develop a reverse-phase HPLC technique. Acetonitrile (ACN) and 0.1% triethylamine (Et3N), pH 2.5, were used as a mobile phase in a 20:80 volumetric ratio to achieve separation with 1.0ml/min, at retention times of about 2.25, 3.02 and 4.38min for glycopyrronium, vilanterol and fluticasone furoate, respectively, with detection at 250nm. Validation in compliance with ICH guidelines verified adequate accuracy, precision, linearity, specificity and robustness, and forced degradation findings confirmed that the method was stability-indicating and sufficient for pharmacopeial quality control of the marketed formulation. Green assessment using the AGREE tool yielded a score of 0.58, supporting its routine use for quality control and stability evaluation of this triple combination.
INTRODUCTION:
Glycopyrronium (GLY) has a molecular weight of 318.4 grams per mole and an IUPAC identification of (1,1-dimethylpyrrolidin-1-ium-3-yl)2-cyclopentyl-2-hydroxy-2-phenylacetate1 with Molecular Formula: C₁₉H₂₈NO₃⁺. GLY is a long-acting muscarinic antagonist that blocks M3 receptors on bronchial smooth muscle2-4. GLY maintains Prolonged airway dilation by blocking acetylcholine-mediated bronchoconstriction, thereby increasing pulmonary airflow and decreasing airway resistance in individuals with chronic obstructive pulmonary disease.
Vilanterol (VIL) is a β₂-adrenergic
receptor agonist with an ultra-long-acting pharmacological profile, activating
selectively β₂ receptors in smooth airway muscle to induce
relaxation5. Vilanterol’s chemical name is
44-[(1R)-2-[6-[2-[(2,6-dichlorophenyl)methoxy]ethoxy]
hexylamino]-1-hydroxyethyl]-2-(hydroxymethyl)phenol, with chemical formula
(C₂₄H₃₃Cl₂NO₅) and molecular weight (486.4
grams per mole)6. It offers a once-daily dosage using this
mechanism, which improves ventilatory function, prolongs bronchodilation, and
allows for the management of chronic respiratory problems.
Fluticasone furoate (FF) is a potent inhaled corticosteroid7-8. The chemical name of FF is [(6S,8S,9R,10S,11S,13S,14S,16R,17R)-6,9-difluoro-17 (fluoromethylsulfanylcarbonyl)-11-hydroxy-10,13,16-trimethyl-3-oxo-6,7,8,11,12,14,15,16 octahydrocyclopenta[a]phenanthren-17-yl] furan-2-carboxylate, with molecular formula C27H29F₃O6S and its molecular weight is 538.6 grams per mole9. FF inhibits the release of inflammatory mediators in the respiratory tract10, stabilises bronchial airways in the long term, and alleviates symptoms resulting from inflammation, such as nasal congestion and excessive mucus production.
Fluticasone Furoate
Figure 1: The Chemical Structures of GLY, VIL and FF
In chronic respiratory conditions, the fixed combination of GLY, VIL, and FF plays complementary roles in bronchodilation and anti-inflammation11. A fixed combination of each Vilfuro-G capsule contains 50µg of GLY, 25µg of VIL, and 100µg of FF. The simultaneous assay of all three drugs is difficult due to their chromatographic behaviours. Most reported analytical methods focus on binary combinations such as Vilanterol-Fluticasone furoate, Glycopyrronium-Vilanterol, or related dual combinations using RP-HPLC12-16, RP-UPLC17-18, or spectrophotometric19 techniques. Only a limited number of studies have addressed the simultaneous estimation of the triple combination of Glycopyrronium, Vilanterol, and Fluticasone furoate, primarily using UHPLC20 methods. This research aims to develop and validate a straightforward, efficient RP-HPLC technique with comprehensive stability-indicating capability and integrated environmental evaluation.
MATERIAL AND METHODS:
Instruments and Apparatus:
A Waters Alliance e2695 was used for chromatographic analysis coupled with photodiode array detection, managed by Empower 2.0 software, using an Inertsil octadecylsilane-bonded C18 column having a length of 250mm, an internal diameter of 4.6mm, and a particle size of 5µm. Buffer pH measurements were made using an Eutech pH 700meter. A Sartorius BSA224S-CW analytical balance was used to weigh the samples and standards. Volumes were measured using Class A Borosil glassware. Sonication of the samples during dissolution and mobile phase degassing was accomplished using a Unichrome UCA-701 ultrasonic bath. The mobile phase was delivered in an isocratic mode by the HPLC system's pump assembly.
Chemicals and Reagents:
GLY, VIL and FF working standards were obtained free as gift samples from a pharma company. HPLC-grade ACN and AR-grade Et3N and OPA (Merck, India) were used. Milli-Q Water was utilised throughout this study, and ACN was used as a diluent for all preparations. Vilfuro-G Capsules, produced by Lupin Ltd. and labelled with 25µg VIL, 100µg FF, and 50µg GLY per capsule, with storage conditions below 25°C, were the marketed formulation utilised in this study.
Preparation of 0.1% Et3N Solution:
0.1% Et3N was prepared by diluting 1ml of Et3N to 1,000ml of HPLC-grade water, followed by pH adjustment to 2.5 using OPA.
Mobile Phase Preparation:
Acetonitrile and 0.1% Et3N were combined in a 20:80 volumetric ratio to prepare the mobile phase. Before usage, this mixture was passed through a 0.45µm PVDF membrane filter.
Standard Solution Preparation:
10mg of GLY, 5mg of VIL, and 20mg of FF were carefully transferred into a 10ml volumetric flask, sonicated to dissolve them in ACN, and then made up to volume. A final standard solution of 10µg/mL of GLY, 5 µg/mL of VIL, and 20µg/mL of FF was prepared by diluting 1ml of the above primary stock solution to 10ml with ACN, then diluting another 1ml of this solution to 10ml using ACN.
Sample Solution Preparation:
Weighed the Contents of 20 capsules accurately. A 5 ml volumetric flask was filled with the sample equivalent to 50µg GLY, 25µg VIL, and 100µg FF. The sample was then dissolved in ACN by sonication for 30minutes. The sample solution was then made up to volume with ACN to yield 10µg/ml GLY, 5µg/ml VIL, and 20µg/ml FF.
Method Development and Optimisation:
By gradually changing the column, ratio of mobile phase, Flow rate, pH, and the detection of wavelength, a method was developed and optimised to obtain clear resolution of GLY, VIL and FF with acceptable peak shape and system performance. A Validation procedure was conducted by utilising the conditions below that produced well-resolved peaks with an optimal analysis time.
A Waters Alliance e2695 was used for chromatographic analysis coupled with photodiode array detection. Separation was carried out using an Inertsil octadecylsilane-bonded C18 column having a length of 250mm, an internal diameter of 4.6mm, and a particle size of 5µm. ACN and 0.1% Et3N buffer (20:80% v/v, pH modified to 2.5 with OPA) were used as the mobile phase, which was pumped at 1.0ml per minute, having a 10µl injection volume, a 6min total run time at room temperature, and detected at 250nm. Mobile phase and all solutions have been passed through the 0.45µm membranes and sonicated prior to analysis to eliminate dissolved gases.
System Suitability:
System suitability was assessed before sample analysis to confirm adequate chromatographic performance. The system suitability assessment findings showed the acceptance requirements (tailing factors <2.0, USP Plate Count ≥2000 for each analyte, and peak resolution >2.0) were satisfied. Consequently, this proved that the system suitability of the chromatography evaluation procedure was satisfactory.
Assay:
The amounts of GLY, VIL, and FF in the Vilfuro-G capsule were measured using RP-HPLC under optimised conditions. After injecting the sample solution, each analyte's percentage was calculated by comparing its area of the peak to the corresponding standard. All values are compared against the pharmacopeial acceptance limits of 98.0 to 102.0%, ensuring that the formulation meets quality standards.
Method Validation:
The proposed RP-HPLC technique for GLY, VIL, and FF was systematically validated using forced degradation tests and evaluation of major analytical performance characteristics in accordance with ICH criteria.
Specificity:
By injecting blank, placebo, standard, and sample solutions, specificity was evaluated. There was no interference from the diluents and excipients if the blank and placebo chromatograms generated no peaks at the GLY, VIL, or FF retention time periods. If the sample peaks matched the standard peaks in both retention duration and spectral purity, the method's specificity for all three analytes was confirmed.
Linearity:
To check whether the RP-HPLC method is linear, calibration curves were constructed using six standard concentrations of GLY(2.5-15µg/.ml), VIL(1.25-7.5 µg/ml), and FF(5-30µg/ml). Peak areas were measured after all concentration levels were introduced into the chromatographic apparatus. The R˛ value was obtained, which is a correlation coefficient used to assess linearity between the concentration of the sample and the peak area of the chromatogram.
Accuracy:
Accuracy was evaluated by standard addition at 80%, 100%, and 120% of the target assay concentration for each drug. Pre-analysed sample equivalent to the label claim was placed in three 5ml flasks, spiked with 0.8, 1.0, and 1.2ml of the mixed standard stock, diluted to the volume of the flask with diluent, sonicated and passed through a 0.45µm membrane and analysed under the optimised conditions. A method was considered accurate if the average recovery of each drug from each level was 98% - 102%.
Precision:
Precision of the RP‑HPLC method was assessed with 3 types: system precision, method precision, and intermediate precision by injecting six injections of the working standards (10µg/ml GLY, 5µg/ml VIL and 20 µg/ml FF). The method is considered highly precise when the percentage RSD of each sample's peak area is less than 2%.
Robustness:
The effectiveness of the RP-HPLC methodology was evaluated based on slight changes in the rate of flow± 0.1ml/min and organic phase composition of the mobile phase ±2% using standard solutions of GLY(10µg/mL), VIL (5µg/mL), and FF (20µg/mL). If the %RSD for the modified chromatography variables was less than 2%, a sign that RP-HPLC methodology produces reliable and repeatable results despite the introduction of minor, intentional changes in the analytical procedure.
Limit of Detection (LOD) and Limit of Quantification (LOQ):
The LOD represents the minimum detectable concentration of each drug that produces a signal-to-noise (S/N) ratio of approximately 3:1, whereas the LOQ represents the minimum concentration that can be quantitatively determined with acceptable precision and accuracy, producing An S/N ratio of approximately 10:1 in the RP-HPLC system.
Forced Degradation Studies:
To evaluate the stability-indicating capability of the method and to ensure reliable quantification of analytes in the presence of their degradation products in the proposed RP-HPLC method, forced degradation studies were conducted. While conducting various degradation studies, 52milligrams of a combined sample containing GLY, VIL & FF was placed in a 5ml vacuum flask, which is considered as sample. For acid degradation, 1 millilitre of 1N HCl was added, and the vacuum flask was maintained at 60 degrees Celsius for 20 minutes and subsequently neutralised with 1N NaOH and diluted with a solvent to produce a volume of 5ml. For alkaline degradation, the same method was used, but instead of HCl, 1N NaOH was used for degradation, and the neutralising solution (1N HCl) was used. For thermal degradation, a standard drug mixture was placed at 105 degrees Celsius for three hours and then diluted using a diluent. For oxidative degradation, the sample was exposed to 1ml of 10% w/v H2O2 and diluted with a solvent to produce a volume of 5ml and placed at 60 degrees Celsius for 20 minutes, followed by a 15minute equilibration time at room temperature. We replaced H2O2 with 10% sodium bisulphite and with 1ml of HPLC-grade water for the reduction and hydrolytic degradation studies, respectively. To conduct the photolytic degradation, the sample was stored in a photostability chamber for 3 hrs and then diluted with the solvent. Following these various studies, the ultimate solutions of each study were filtered through a PVDF membrane filter of 0.45µm and injected into an HPLC apparatus. The method was considered stability-indicating if the peak purity angle for each analyte was lower than the corresponding purity threshold under all stress conditions.
RESULTS AND DISCUSSION:
GLY, VIL, and FF were simultaneously analysed using the optimised chromatographic conditions. Performance was evaluated on the method suitability and method validation parameters supported by the International Council for Harmonisation. Results obtained from this optimisation and validation are discussed below.
System Suitability Findings and Optimised Chromatogram:
Table 1: Findings of system suitability for GLY, VIL and FF
|
S. No |
Name |
Retention Time |
Peak Area |
USP Resolution |
USP Tailing |
USP Plate Count |
|
1 |
GLY |
2.245 |
1560207 |
-- |
1.05 |
8916 |
|
2 |
VIL |
3.018 |
772146 |
3.71 |
1.12 |
6057 |
|
3 |
FF |
4.376 |
3084721 |
6.92 |
0.94 |
11491 |
Figure 2: Optimised RP-HPLC chromatogram of GLY, VIL and FF
The optimised chromatographic conditions outlined in the materials and methods part were used to separate GLY, VIL, and FF. Under these conditions, all three analytes produced sharp, symmetrical peaks, and system‑suitability parameters are within acceptable limits. System suitability outcomes and optimised chromatogram are displayed in Table 1 and Figure 2, respectively.
Assay:
Table 2: Findings of Assay for GLY, VIL and FF
|
Parameter |
GLY |
VIL |
FF |
|
Brand |
VILFURO-G |
VILFURO-G |
VILFURO-G |
|
Peak Area–Injection 1 |
1,563,102 |
776,108 |
3,056,271 |
|
Peak Area–Injection 2 |
1,583,218 |
775,912 |
3,065,546 |
|
Average Sample Area (n = 2) |
1,573,160 |
776,010 |
3,065,546 |
|
Label Amount (mg) |
0.050 |
0.025 |
0.100 |
|
Standard Purity (%) |
99.7 |
99.8 |
99.9 |
|
Amount Found (mg) |
0.050 |
0.030 |
0.099 |
|
% Assay |
100.0 |
100.2 |
99.7 |
Assay results are shown in Table 2. These findings demonstrate the utility of the RP-HPLC as an acceptable and reliable routine assay method to assist in providing proof of accuracy of GLY, VIL and FF without interferences due to formulation excipients.
Specificity:
Blank Chromatogram
Placebo Chromatogram
Figure 3: Blank and Placebo chromatograms
Figure 3 gives the chromatograms of the blank and placebo. Specificity of this method was demonstrated by the absence of interfering peaks from blank and placebo solutions at the retention times of GLY, VIL, and FF.
Linearity:
Figure 4: Linearity outcomes for GLY, VIL and FF
As the concentration increased, a gradual increase in peak area was seen for all three analytes, confirming a linear relation between concentration and response, as evidenced by the regression equations with R˛ values equal to 0.9998 for GLY, 0.9992 for VIL and 0.9996 for FF, as shown in Figure 4.
Accuracy:
Table 3: Accuracy Outcomes for GLY, VIL, FF
|
Drug |
Accuracy Level (%) |
Average Peak Area |
Amount Spiked (mg) |
Average Amount Found (mg), %RSD (n=3) |
Mean % Recovery (n=3) |
|
GLY |
80 |
2827331 |
0.09 |
0.0903, 0.64 |
100.4 |
|
100 |
3140668 |
0.10 |
0.1000, 1.00 |
100.0 |
|
|
120 |
3440192 |
0.11 |
0.1100, 0.91 |
100.0 |
|
|
VIL |
80 |
1390870 |
0.045 |
0.0433, 13.32 |
99.8 |
|
100 |
1556429 |
0.050 |
0.0500, 0.00 |
100.0 |
|
|
120 |
1713558 |
0.055 |
0.0567, 10.19 |
100.5 |
|
|
FF |
80 |
5561562 |
0.18 |
0.1810, 0.55 |
100.6 |
|
100 |
6158341 |
0.20 |
0.2010, 0.50 |
100.5 |
|
|
120 |
6741120 |
0.22 |
0.2190, 0.46 |
99.5 |
The recovery results for each drug GLY (100.0-100.4), VIL(99.8-100.5) and FF(99.5-100.6) are presented in Table 3, showing that the RP-HPLC method is accurate and unaffected by the excipients used in the Vilfuro-G formulation.
Precision:
Table 4: Precision Results for GLY VIL and FF
|
Precision Type |
Injection No |
GLY Peak Area |
VIL Peak Area |
FF Peak Area |
|
System Precision |
1 |
1560207 |
772146 |
3084721 |
|
2 |
1572103 |
775134 |
3061846 |
|
|
3 |
1578347 |
778130 |
3074612 |
|
|
4 |
1564028 |
772415 |
3094620 |
|
|
5 |
1569409 |
776412 |
3064158 |
|
|
6 |
1571310 |
774311 |
3072814 |
|
|
Mean |
1569234 |
774758 |
3075462 |
|
|
%RSD |
0.41 |
0.30 |
0.40 |
|
|
Method Precision |
1 |
1568217 |
771462 |
3062718 |
|
2 |
1559241 |
770204 |
3098547 |
|
|
3 |
1574671 |
772521 |
3042719 |
|
|
4 |
1572628 |
778234 |
3051614 |
|
|
5 |
1563041 |
779328 |
3051987 |
|
|
6 |
1571032 |
773009 |
3062154 |
|
|
Mean |
1568138 |
774126 |
3061623 |
|
|
%RSD |
0.38 |
0.48 |
0.64 |
|
|
Intermediate Precision |
1 |
1568720 |
770381 |
3084627 |
|
2 |
1558287 |
771208 |
3046170 |
|
|
3 |
1567341 |
770341 |
3096127 |
|
|
4 |
1589125 |
778102 |
3061278 |
|
|
5 |
1591813 |
773609 |
3054681 |
|
|
6 |
1569646 |
771128 |
3078213 |
|
|
Mean |
1574155 |
772462 |
3070183 |
|
|
%RSD |
0.84 |
0.39 |
0.63 |
The precision studies for GLY, VIL and FF are included in Table 4. The statistical analysis of the Peak Area measurements shows that %RSD for each study was below 2%, indicating that the RP-HPLC Analytical Procedure was suitable for concurrent estimation of the three Compounds.
Robustness:
The robustness of the proposed RP-HPLC technique was confirmed by the fact that adjustments in flow rate (±0.1 ml/min) and organic phase composition (±2%) had no significant impact on retention duration, peak shape, column efficiency, and analytical response for GLY, VIL, and FF.
Limit of Detection (LOD) and Limit of Quantification (LOQ):
The LOD and LOQ values for GLY, VIL, and FF were 0.30 and 1.00 µg/mL, 0.15 and 0.50 µg/mL, and 0.60 and 2.00 µg/mL, respectively, based on S/N ratios of 3:1 and 10:1, confirming the adequate sensitivity of the method.
Forced Degradation Studies:
Table 5: Forced Degradation results for GLY, VIL and FF
|
Drug |
Condition |
Peak Area |
Assay (%) |
Degrada-tion (%) |
Peak Purity Angle |
Peak Purity Threshold |
|
GLY |
Control |
1568236 |
100 |
0 |
2.143 |
4.055 |
|
Acid |
1543284 |
98.4 |
1.6 |
2.156 |
4.069 |
|
|
Alkali |
1401287 |
89.3 |
10.7 |
2.121 |
4.027 |
|
|
Peroxide |
1376241 |
87.7 |
12.3 |
2.169 |
4.019 |
|
|
Reduction |
1532640 |
97.7 |
2.3 |
2.108 |
4.011 |
|
|
Hydrolysis |
1552304 |
98.9 |
1.1 |
2.175 |
4.003 |
|
|
Thermal |
1391647 |
88.7 |
11.3 |
2.164 |
4.009 |
|
|
Photolytic |
1540364 |
98.2 |
1.8 |
2.138 |
4.012 |
|
|
VIL |
Control |
774732 |
100 |
0 |
1.423 |
3.966 |
|
Acid |
700914 |
90.5 |
9.5 |
1.482 |
3.912 |
|
|
Alkali |
741504 |
95.7 |
4.3 |
1.416 |
3.946 |
|
|
Peroxide |
689471 |
89.0 |
11.0 |
1.498 |
3.937 |
|
|
Reduction |
761349 |
98.3 |
1.7 |
1.472 |
3.984 |
|
|
Hydrolysis |
747214 |
96.4 |
3.6 |
1.405 |
3.921 |
|
|
Thermal |
759546 |
98.0 |
2.0 |
1.468 |
3.965 |
|
|
Photolytic |
756281 |
97.6 |
2.4 |
1.473 |
3.914 |
|
|
FF |
Control |
3076176 |
100 |
0 |
3.625 |
5.298 |
|
Acid |
2727454 |
88.7 |
11.3 |
3.617 |
5.212 |
|
|
Alkali |
2751374 |
89.4 |
10.6 |
3.619 |
5.241 |
|
|
Peroxide |
2613842 |
85.0 |
15.0 |
3.688 |
5.236 |
|
|
Reduction |
2994271 |
97.3 |
2.7 |
3.636 |
5.279 |
|
|
Hydrolysis |
2941870 |
95.6 |
4.4 |
3.607 |
5.245 |
|
|
Thermal |
2631745 |
85.6 |
14.4 |
3.618 |
5.261 |
|
|
Photolytic |
3059187 |
99.5 |
0.5 |
3.604 |
5.258 |
In all stress conditions, all three drugs GLY, VIL, and FF demonstrated controlled degradation, while no change occurred in the control sample, and the peak purity angle for each analyte was found to be lower than the corresponding purity threshold, as shown in Table 5, confirming the stability-indicating nature of the method.
Green Analytical Assessment:
The RP-HPLC procedure for GLY, VIL, and FF received a score of 0.58 on the AGREE (Analytical greenness) assessment method, which gives the environmental consciousness in relation to the twelve rules of Green Analytical Chemistry. An AGREE score higher than 0.50 denotes a moderately green, acceptable environmental analytical methodology, according to the above recent research in green analytical chemistry. An AGREE Circular Diagram (Figure 5) shows the entire greenness performance of the RP-HPLC method designed to obtain a combined estimation of GLY, VIL, and FF.
Figure 5: Score of Analytical greenness
CONCLUSION:
This Reversed-Phase High-Performance Liquid Chromatography method is a straightforward and reproducible way to measure three active combined ingredients (GLY, VIL and FF) in a single inhalation aerosol formulation. After optimisation of chromatographic conditions, analytes were adequately separated with good retention, good resolution, and symmetry of peaks. Additionally, the analytical methods meet ICH requirements through various studies. Furthermore, forced degradation studies demonstrate that this analytical procedure is suitable for a stability-indicating assay. The AGREE assessment rating of 0.58 indicates that the method is moderately green. As a result, the analytical approach will be helpful for routine stability testing and quality control of multi-drug inhalation products.
REFERENCES:
1. PubChem Compound Summary for CID 3494, Glycopyrronium. National Center for Biotechnology Information. Retrieved December 15, 2025. Available on https://pubchem.ncbi.nlm.nih.gov/compound/Glycopyrronium.
2. Adimoolam BM, Jangam A, Dadinaboyina SB, Ragampeta S, Thota JR. Stability studies of glycopyrronium bromide and in vitro toxicity evaluation of its major transformation products by LC-MS/MS and NMR. European Journal of Mass Spectrometry. 2025 May 4:14690667251339714.
3. Ridolo E, Milanese M, Barone A, Nicoletta F, Ottoni M, Cosini FF, Passalacqua G, Lombardi C. Extrafine formulation of beclomethasone dipropionate/formoterol fumarate/glycopyrronium bromide delivered via pressurized metered-dose inhaler in the treatment of asthma: a review. Therapeutic Advances in Respiratory Disease. 2025 Apr; 19:17534666251332076.
4. Patel K, Patel D, Dwivedi J, Newal P, Meshram D. A review on analytical techniques for estimation of anti cholinergic drug: Glycopyrronium bromide. Asian Journal of Pharmaceutical Analysis. 2017 Dec 24; 7(4):239-42.
5. Jin LU. Adrenergic agonist therapy and emerging acupuncture targets in asthma: A review of mechanisms and therapeutic implications. World Journal of Acupuncture-Moxibustion. 2025 Aug 5.
6. PubChem Compound Summary for CID 10184665, Vilanterol. National Center for Biotechnology Information. Retrieved December 15, 2025. Available on https://pubchem.ncbi.nlm.nih.gov/compound/Vilanterol.
7. Daley-Yates PT, Deans A, Mehta R, Sousa AR. Comparative clinical pharmacology of mometasone furoate, fluticasone propionate and fluticasone furoate. Pulmonary Pharmacology & Therapeutics. 2022 Dec 1; 77: 102171.
8. Murali M, Kothandan S, Chandrasekaran M, Singh SK, Kuppusamy G, Radhakrishnan A. Fluticasone and Cyclophosphamide towards the track of personalized Medicine. Research Journal of Pharmacy and Technology. 2021; 14(1): 179-84.
9. PubChem Compound Summary for CID 9854489, Fluticasone Furoate. National Center for Biotechnology Information. Retrieved December 15, 2025. Available on https://pubchem.ncbi.nlm.nih.gov/compound/Fluticasone-Furoate.
10. Golla R, Kumar K. Current Status of Ultra Long-Acting Beta Agonists in Children. Indian Pediatrics. 2025 Jul 23:1-4.
11. Patel C, Sahoo D, Sheth V, Jain MK, Koppula R, Verma S, Kumar D, Panda JK, Parmar D, Kansagra K, Pathak H. A Novel Metered Dose Inhaler Formulation of Triple-Drug Fixed-Dose Combination of Vilanterol, Glycopyrronium, and Fluticasone Furoate: A Phase III, Randomized, Multicenter Trial to Evaluate the Efficacy and Safety in Indian Patients with Uncontrolled Asthma. Clinical Therapeutics. 2025 Oct 29.
12. NB L, AG G, NB B, SB B. RP-HPLC Method for Simultaneous Quantification of Vilanterol and Fluticasone Furoate: Development and Validation. Eurasian Journal of Analytical Chemistry. 2024 Jan 1; 19(1).
13. Masimukku SK, Chintala R. Development and validation of stability indicating RP-HPLC method for fluticasone furoate and vilanterol in pharmaceutical formulations. Indian Drugs. 2018 Mar 1; 55(3).
14. Patel D, Namdev KK, Verma K, Gururani R, Tiwari A, Kumar P, Dewangan RP, Wabaidur SM, Sharma S, Dwivedi J. HPLC-UV and spectrofluorimetric methods for simultaneous estimation of fluticasone furoate and vilanterol in rabbit plasma: A pharmacokinetic study. Journal of Chromatography B. 2019 Nov 15; 1132:121842.
15. Kulkarni SV, Gulave VR, Musmade BD, Bhope SG, Nagar M, Panchgalle SP, More VS. Impurity Profiling Method Development, Validation, and Relative Response Factor Estimation for Fluticasone Furoate and Vilanterol Trifenatate Dry Powder Inhaler Using RP-HPLC with UV/PDA Detector. Pharmaceutical Chemistry Journal. 2025 Aug; 59(5):598-610.
16. Babu GR, Sowjanya M, Ramayyappa M, Karishma MA, Bharathi V, Swathi GR, Renuka C, Krishna GG, Lavanya M. Rapid RP-HPLC Method for Simultaneous estimation of Umeclidinium and Vilanterol in human plasma. Research Journal of Pharmacy and Technology. 2023 Mar 1; 16(3):1017-22.
17. Babu PV, Devi DA. Stability Indicating Assay Method Development and Validation by RP-UPLC with PDA detector for Simultaneous Estimation of Glycopyrrolate and Neostigmine in Pharmaceutical dosage form. Research Journal of Pharmacy and Technology. 2023 Jul 1; 16(7):3219-22.
18. REDDY BV, Vejendla A, Dittakavi R, REDDY SN. Development and validation of simultaneous estimation of glycopyrrolate and formoterol fumarate in its bulk and pharmaceutical rota caps dosage form by using RP-UPLC. International Journal of Applied Chemistry. 2020; 7(3):49-56.
19. Masimukku SK, Chintal R. Development and validation of spectrophotometric methods for simultaneous estimation of vilanterol and fluticasone furoate in pharmaceutical formulations. Asian J Pharm Clin Res [Internet]. 2017 Apr. 1 [cited 2026 Mar. 1]; 10(4): 302-5. Available from: https://journals.innovareacademics.in/index.php/ajpcr/article/view/16748
20. Anitha L, Beda DP. Implementation of quality by design and green chemistry principles to validate the stability indicating method on UHPLC for the simultaneous quantification of Glycopyrronium, Fluticasone, Vilanterol in inhaled formulation. Green Analytical Chemistry. 2024 Dec 1; 11:100160.
|
Received on 31.01.2026 Revised on 01.04.2026 Accepted on 16.05.2026 Published on 10.10.2026 Available online from October 17, 2026 Asian J. Pharm. Res. 2026; 16(4):371-377. DOI: 10.52711/2231-5691.2026.00054 ©Asian Pharma Press All Right Reserved
|
|
|
This work is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International License. Creative Commons License. |
|